Quick Detail
Place of Origin
Brand Name
Mitech
Model Number
MH310
HS-CODE
85-
Package & Delivery Lead Time
Package
ABS instrument case in the carton box
Delivery Lead Time
1-2 working days
Detailed Description
Mitech MH660 Leeb Hardness Tester support English different from ultrasonic hardness tester
Technical Parameters
- - - -
Measuring range | 170 to 960 HLD |
Display | color TFT LCD,320240 dots, adjustable backlight |
Communication | Blue tooth/USB1.1 |
Weight | Approximately 220g |
Dimensions | 120mm67mm31mm |
Integrated data memory | 500 measurement series |
- - - -
Application
Parts of the Main body
1,The main unit 2,Keypad 3,LCD display
4, Battery compartment 5,Socket of impact device 6, Bluetooth status
7,Impact device 8, Label 9, USB port
Screen Display
Material Selected material.
Impact device Type of impact device connected to the main unit
Impact direction Impact direction setting (AUTO when automatically detecting the direction and auto correction); Or detected impact direction after impact in auto direction mode.
File name Next file name to store the measurement series.
USB status Appears when the USB connection succeeds.
Time Real time clock of the instrument system in HH:MM format.
Battery capacity Shows the rest capacity of the batteries.
Measurements count Number of measurements in a series.
Average times Set number of measurements per series.
Test value Shows last individual measured value before the test series closes; Or shows the average measured value when the test series closes.
Scale Hardness/Strength scale to display the test value.
Statistics Shows minimum, maximum and average value of the test series.
Functions Functions that can be executed. Functions are programmed and assigned to the function keys F1 to F3. Note that the assignment of the function keys changes depending on the current dialog/menu.
Testing operation could be carried out under this main screen. After each impact, it will display a singled measured value; impact count plus one; the buzzer would alert a long howl if the measured value is not within the tolerance limit. When reaching the average times, the buzzer will alert a long howl. After 1 to 2 seconds, the buzzer will alert a short howl, and display the average measured value.