Quick Detail
Place of Origin
HS-CODE
S04-
Package & Delivery Lead Time
Detailed Description
DX-2012H DC Hysteresis Graph Test System
Automatic measurement of the demagnetization curve of permanent magnetic material such as ferrite, AlNiCo, NdFeB, SmCo, etc. Accurate measurement of the magnetic characteristic parameters of remanence Br, coercive force HcB, intrinsic coercive force HcJ and maximum magnetic energy product (BH)max.
Windows measurement software applied simply. The product conforms to China National Standards GB/T 3217-92, GB 11209-89 and international standard IEC 404-5.
Adopt ATS structure, users can customize different configuration as required: According to the intrinsic and size of measured sample to decide the electromagnetic size and corresponding testing power supply; Select different measuring coil and probe according to the option of measuring method. Decide to whether choosing fixture in accordance with sample shape. Decide to whether selecting heating installation or not according to the test requirement.
General Features of DC Hysteresis Graph (DX-2012H)
Automatic measurement on saturation hysteresis loop and magnetic characteristic parameters of generally permanent-magnet material such as Ferrite, Rubber & Plastic Magnet and AlNiCo, etc.: Js, Br, HcB, HcJ, (BH)max and µrec.
Automatic measurement on demagnetization curve and magnetic characteristic parameters of rare earth permanent-magnet materials such as NdFeB and SmCo, etc. at the second quadrant: Br, HcB, HcJ, Hk(H90) and (BH)max.
Test sample shapes: circular ring, round cake, square, tile and other irregular shapes.
Adopt B coil + fluxmeter to measure magnetic induction, zero shift of integrator can be self-corrected through software.
Adopt J coil + fluxmeter to measure magnetical polarization, remnant coil area of J coil can be automatically compensated through software.
Magnetic field intensity can be measured with Hall magnetometer, nonlinear error of hall probe can be corrected through software, within 0 ~ 2.4T range, nonlinear error can be controlled within±0.2%.
Adopt H coil + fluxmeter to measure magnetic field intensity, Hall magnetometer only used to indicate zero point of magnetic field so as to thoroughly eliminate nonlinear error of hall probe.
Optimal range of field voltage, fluxmeter and magnetometer can be automatically set up.
Magnetization, testing and demagnetization of general permanent magnetic sample completed at one time, time set up 20 seconds ~ 60 seconds.
Rare earth permanent magnetic samples need saturation magnetizing before testing, testing time 60 seconds ~ 120 seconds.
The sample after tested is in demagnetization or magnetizing status, freely selected by users.
Select heating head and temperature controller to detect magnetic characteristics of Ferrite, AlNiCo and rare earth material under high temperature (maximal 220°C).
Select shoe jig for direct measurement on magnetic shoe.
Technical Data of DC Hysteresis Graph (DX-2012H)
System Specifications
Test AlNiCo standard sample, measuring error on entire magnetic hysteresis cycle estimated: B: 1%, H: 1%
Parameters measured Br(%) HcB(%) HcJ (%) (BH)max(%)
Uncertainty (k=2) 1 1 1 2
Repeatability (constant temperature) ± 0.3 ± 0.3 ± 0.3 ±1
Instrument Specifications
TPS-2000H Test Power Supply DCT150 Electromagnet
Output Voltage: 0 ~±100V
Output Current:±20A (maximum value)
Stability: superior to 0.05%
Stability: superior to 0.05% Head Diameter: 150mm shrink to 80mm
Head Spacing: 0 ~ 50mm (continual adjustable)
Maximal Magnetic Field: 2.4T (head spacing 10mm, coil current 30A)
LZ-210B Fluxmeter DCT200 Electromagnet
Range: 2.5, 5, 10, 20mWb
Full Range Precision: 0.5%
Zero Shift: 2µWb /30 seconds (2.5mWb) Head Diameter: 200mm shrink to 120mm
Head Spacing: 0 ~ 50mm (continual adjustable)
Maximal Magnetic Field: 2.4T (head spacing 10mm, coil current 40A)
LZ-210H Magnetometer DCT250 Electromagnet
Range: 0.5, 1, 1.5, 3T
Full Range Precision: 0.5%
Non-Linearity:±0.2% (software correction) Head Diameter: 250mm shrink to 150mm
Head Spacing: 0 ~ 50mm (continual adjustable)
Maximal Magnetic Field: 2.7T (head spacing 10mm, coil current 80A)
DCT130 Electromagnet PC6616 A/D D/A card
Head Diameter: 130mm
Head Spacing: 0 ~ 50mm (continual adjustable)
Maximal Magnetic Field: 1.5T (head spacing 10mm, coil current 20A) A/D Conversion: 16 bits,16 channels
Conversion Time: < 10µs
Conversion Accuracy: 0.02% (full range)
D/A Conversion: 12 bits, 2 channels
Conversion Accuracy: 0.1% (full range)
Product Family of DC Hysteresis Graph (DX-2012H)
Standard Equipment
Serial No. Product Model DX-2012H130 DX-2012H150 DX-2012H200 DX-2012H250
1 Electromagnet Model DCT130 DCT150 DCT200 DCT250
2 Test Power Supply 2kW 3kW 4kW 8kW
3 Fluxmeter Configuration One LZ-210B Two LZ-210B
4 Magnetometer LZ-210H
5 Data Acquisition Card PC6616
6 Measurement Software HMTest partial functions HMTest all functions
7 Computer andPrinter Dell Dimension E-520n computer+ HP1020 laser printer
8 Standard Specimen One AlNiCo standard sample
9 Fixed Coil One B coil One J coil J coil + H coil J coil + H coil
10 Test Specimen Category AlNiCo
Ferrite AlNiCo
Ferrite
NdFeB AlNiCo
Ferrite
NdFeB
SmCo AlNiCo
Ferrite
NdFeB
SmCo
11 Specimen Size
(with round sample as an example) Diameter 3 ~ 100mm
Thickness 3 ~ 50mm
< 130 - 1.2H General permanent magnetic:
3 ~ 60mm
H3 ~ 50mm
Rare earth permanent magnetic:
5 ~ 20mm
H5 ~ 15mm
General permanent magnetic:
3 ~ 90mm
H3 ~ 50mm
Rare earth permanent magnetic:
5 ~ 30mm
H5 ~ 20mm
General permanent magnetic:
3 ~ 120mm
H3 ~ 50mm
Rare earth permanent magnetic:
5 ~ 50mm