Quick Detail
Place of Origin
HS-CODE
90-
Package & Delivery Lead Time
Detailed Description
Features:
12-lead simultaneous standard resting ECG
One-touch operation
Concept of floating input circuit meets IEC safety standard
High-resolution thermal printing array system
Frequency response is as high as 150Hz
Capable of printing continuously on channel trace and annotations including lead mark and parameter such as sensitivity, paper speed and filter operation status
High resolution digital filter inhibits baseline drift without affecting ECG waveform
Rechargeable battery supporting more than 50 patient examinations
Safety standard
IEC class I, type CF
Input circuit
Floating; Protection against Defibrillator effect
Lead
Standard 12 Leads, Lead change automatically
Patient current leakage
<10μA
Input impedance
=>10ΜΩ
Calibrating voltage
1mV+_3%
A/D conversion
12 bit
Frequency response
0.05Hz~150Hz (IEC)
Time constant
>3.2seconds
CMRR
>80dB
>100dB (with filter)
EMG filter
35Hz (-3dB) / 25Hz (-3dB)
Sensitivity
1/2, 1, 2 (cm/mV), Conversion Deviation=<5%
Recording system
Thermal-array
8 dots/mm ( vertical )
16 dots/mm ( horizontal, 25mm/sec )
Paper speed
25, 50mm/s+_3%
Recording paper
50mm,20m, high-speed roll
Input circuit current
=<50nA
Power supply
AC: 220V / 110V, 50/60Hz
DC: 12V rechargeable battery
Skin voltage tolerance
+_300mV
Noise level
<15uVp-p
Dimensions
310X230X70mm
Dimension of packing
400X280X160mm
Weight
2.75kg