Quick Detail
Place of Origin
HS-CODE
90-
Package & Delivery Lead Time
Detailed Description
Infusion Pump Technical Specification
- Bolus: 600ml/h
- Flow rate range: 1-600ml/h
- Volume limit: 1-9999 ml
- Accuracy:±5%
- KVO flow rate:1ml/h, keep vein open KVO rate
- Power supply: AC100-240V, 50/60 Hz, 25VA
- Water Proof: IP *I
- Battery: Rechargeable lithium polymer battery,7.4 V,1650mAh
- Electrical safety: compliance with the requirements of IEC 60601-1.
- Max. Power consummation: 25W
- Battery recharge: When the pump is connected to the AC power, the battery will automatically
recharge About 8-14 hours to recharge fully Can run for more than 5 hours continuously
after fully recharged.
- Fuse Type:220V 2A*2,12V 2A*2
- Display or information: Flow rate, volume limit, accumulated volume, power indicator light, bed
No., air, occlusion, empty.
- Alarm function: Infusion completion, occlusion, air bubble, low battery, control abnormal, no
AC power supply, installation error.
- Max. size of outer shell:120*140*190 mm-length* width* height
- Max. weight:<2.5 kg
- Classification: Type B.
- Outer shell material: ABS plastic
- operating condition: Environment Environment temperature:+5℃- +40℃,atmosphere
pressure 50-106kPa,related humidity 30%-90%
- Storage and transport condition: Environment temperature -15℃- +50℃,atmosphere
pressure 50-106kPa,relative humidity 30%-90%.
- Applicable infusion pipe: All standard infusion pipe use "double dove" to test
- EMC: Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
contec_cherise@yahoo.cn
contec_cherise@hotmail.com