Features
■ Color TFT Screen, waveform up to 8 channels
■ Light and portable with build-in rechargeable battery
■ UP to 10 kinds of monitor parameters
■ Arrhythmia analysis, pace-marker detection and S-T Segmentanalysis
■ 72 hour storage and review of trend grams and tables, 40 second review of olographic
wavefo rms.
■ Built-in recorder(option)
■ All-round monitor of adult ,pediatric and neonatal.
■ Electrosurgical unit and defibrillation protected
■ Standard parameters: ECG, NIBP, SPO2, Respiration, Temperature, Pulse rate
■ Option parameters: EtCO2, CO, IBP, Thermal Printer
ECG
■ 5-lead or 3-lead selectable
■ Input: 5-lead: RA; LA; RL; LL; V or R; L; N; F; C
■ Lead selection: I; II; III; avR; avL; avF; Vx; CAL
■ Gain selection: 12.5mm/s, 25mm/s, 50mm/s
■ Accuracy: ?bpm or ?%, whichever is greater
■ Protection: Withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation
■ Calibration signal: 1mV?%
■ Operation modes: Diagnostic, Monitor, Surgery
■ Alarm range: Adult: 15-300bpm
■ Pediatric/Neonatal: 15-350bpm
■ Alarm range: -2.0mV~2.0mV
■ S-T detection: YES, Arrhythmia analysis: YES
■ Alarm: YES, audible and visual alarm, alarm events recallable
NIBP
■ Method: Oscillometry
■ Operation modes: Manual/Automatic/STAT
■ Measurement unit: mmHg/KPa selectable
■ Measurement types: Systolic, Diastolic, Mean
■ Measurement range: Adult: 10-270mmHg
■ Over-pressure protection: YES
■ Resolution: 1mmHg
■ Accuracy: Mean error and standard deviation per ANSI/AAMI SP-10
■ Alarm: Systolic,Diastolic,Mean
SpO2
■ Measurement range: 0-100%
■ Accuracy: ?digit (70~100%);
■ 0~69% unspecified
■ Alarm range: 0-100%
Respiration
■ Method: RA-LL impedance
■ Pediatric/Neonatal: 6-150rpm
■ Alarm range: Adult: 6-120rpm
■ Pediatric/Neonatal: 6-150rpm
■ Apnea alarm: YES
Temperature
■ Resolution: 0.1℃
■ Channel: Dual-channel
■ Alarm range: 0-50℃
Option:
EtCO2
■ Measurement Mode: Side-stream or mainstream
■ Measurement range: EtCO2: 0-99mmHg
■ Accuracy: CO2 concentration
■ Apnea alarm: YES
CO
■ Method: Thermo-dilution
■ Measurement range: CO: 0.1-20 lit./min.
IBP
■ Channel: 2 channels
■ Pressure transducer: Sensitivity: 5( mV/V/mmHg)
■ Transducer sites: ART, PA, CVP, RAP, LAP, ICP
■ Alarm range: -10~300mmHg