Features:
12.1'' TFT Big-screen, real-color, wide-visual-angle and high-brightness LCD display, Chinese-English language interface;
Can monitor the main waveforms of standard synchronous 7-lead and 12-lead,analyze on ST segment to facilitate diagnosis for doctors, choose the source of three kinds of Heart Rate, have waveform-filling and 5-min cardiac electric replay function, measure synchronous many kinds of standard regular parameters, pulse conducting time and cuff pressure, and save the integrate patient information and review the monitoring-record;
With the formed by network, bed-side machines and central multi-parameter monitor system, constituting the web of monitoring and trend-data display and 256 hours trend review;
With standard six-parameter monitoring, adding single/double IBP, Et CO2(side stream),monitoring fetal function to provide accurate quickening and TOCO;
With manual printing and alarm trigger printing functions available, and able to use together with high-frequency electrical cutter and defibrillator.
Specifications:
1.Safety level
Class I anti-electroshock equipment, ECG(RESP), TEMP--Type CF, SPO2, NIBP--BF
2. Power supply
AC 100-250V , 50/60Hz, Pmax=60VA, FUSE 0.5A
3. Display
True color,12.1'' type LCD, resolution 640x480
4. Signal interface
External display interface: standard VGA
External communication interface: RS-485
External print interface: standard combining-port
5. Recorder
Recording width: 48mm
Paper speed: 25mm/s
Trace: 3 channel
6. ECG specifications:
Lead mode:12 leads (R,L, F, N, C1, C2, C3, C4, C5, C6)
Lead selection: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform:12 channels
Gain: ×0.5, ×1.0, ×2.0
Modes: diagnosis, monitoring, operation
Input impedance:>5MΩ
CMRR: >90dB
Electrode offset potential: ±300mV
Leakage current :<30μA
ECG signal range: ±10mV(Vp-p)
Frequency response (bandwidth):diagnosis 0.05-100Hz
Monitoring: 0.5-40Hz
Operation: 1-20Hz
Measurement range: 10-254bpm
Resolution: 1bpm
7. RESP specifications:
Method: Impedance between RA-LA
Input impedance: >2.5MΩ
Bandwidth: 0.3-2Hz
Resp rate: 0-120rpm
Measuring Range: 0-120rpm
Resolution: 1bpm
8. NIBP specifications:
Method: oscillation
Mode: manual, periodic, continuous
Measuring interval in periodic mode: 3, 10, 45, 60 min
Continuous mode: 5 min
Pulse rate range: 40-240bpm
Overpressure protection: adult 290 mmHg child 145 mmHg
9. SPO2 specifications:
Measuring range: 0-100%
Resolution: 1%
Actualization interval: 1sec
Alarm delay: 10 sec
Pulse rate: measuring range: 0-254bpm
10. TEMP Specifications:
Measuring range: 0-50℃
Resolution: 0.1℃
Actualization interval: 1sec
Average time constant: <10 sec
Packaging collocation
Packaging Dimension: 45cmx45cmx35cm
Gross weight:12kg
Normative collocation: 220V power cable 1pcs, BPM-9500 six-parameter, BPM-9500A six-parameter+ printer, BPM-9500B six-parameter+printer+EtCO2, BPM-9500C six-parameter+printer+single fetal monitor.