Quick Detail
Place of Origin
HS-CODE
90-
Package & Delivery Lead Time
Detailed Description
JPD-600P 12.1 LCD
Features:
1. 12.1 high-resolution color TFT LCD display, support plug-in large-screen CRT display
2.Standard Configure: FHR,TOCO,Fetal Movement, ECG, Spo2, NIBP, RESP, TEMP, Pulse Rate
3. Optional twins
4. Portable, convenient monitor
5. Built-in rechargeable battery, AC-DC power supply, without interruption Monitor
6.Dynamic save and display the Monitoring information within 48 hours
7.Both Use lonely or Connect with a Central Monitor network
8. Support the extension almost for 128 units central monitor system.
9. Quality “Sanyo” speaker take more clear and lound sound.
10. usual thermal paper,more economical.
Specification:
FHR:
Ultrasound Probe: 12pcs chip, Multi-wafer,wave-beam,high sensitive
Ultrasound frequency: 1MHz±5%
Ultrasound intensity: <3mW/cm2
FHR range: 30-240bpm
Resolution:1bpm
Accuracy:± 1bpm/±1%
TOCO:
TOCO range: 0-100 (relative unit)
Resolution:1
Non-linear error:≤±5%
Fetal Movement:
Button Hand Count( Automatic cumulative number of movements)
NIBP:
Measurement Units:mmHg/Kpa optional
Pressure protection: A double security protection software and hardware
Work Mode: Auto,Manual
Measurement intervals: 2-60 minutes set up
Survey: : Systolic blood pressure:60-240 mmHg
Average Blood Pressure:40-200 mmHg
Diastolic Blood Pressure: 30-180 mmHg
Resolution:1mmHg
Accuracy:±3mmHg
Pulse rate:40-240bpm
SPO2:
Range:70%-99%
Resolution:1%
Accuracy:±2 Effective digital
Pulse Rate:
Range:40-240bpm
ECG:
Range: 30-300bpm
Accuracy:±1bpm/±1%
Choice:I、II、 III、aVR、aVL、aVF、V
Scan Speed:12.5mm/s、25mm/s、50mm/s
Gain choice:x0.25、x0.5、x1、x2
Application:Diagnosis,Monitor,Surgery
RESP:
Range:1-99rpm
Resolution:1rpm
Accuracy:±2rpm
TEMP:
Range:26℃- 45℃
Resolution:0.1℃
Accuracy:±0.2%
Work environment:
Operation Temperature: 5℃- 40℃
Relative temperature:≤80%
Atmospheric pressure: (86-106)Kpa,32℃- 45℃:±0.1℃
Printer
50mm usual thermal paper,more economical