Model: CP08GNL-B3
CP08GNL-B3 oxygen analyzer - wave soldering, reflow soldering, brazing dedicated - can be achieved T1-3000TA requirements. The new CP08GNL-B3 (reflow, wave soldering furnace only) oxygen analyzers, using the WAN than testing principle, with the world's advanced technology to break the traditional pattern detection, pioneering the realization of the on-line measurement - no benchmark gases (oxygen concentration from environmental impact), the sensor is interchangeability, and test high-precision calibration cycle length, small size, low power consumption, stable and reliable features.
Instrument features:
Control points set arbitrary output
On the threshold alarm point in the whole range of arbitrary settings.
With 0 ~ 5 V. DC or 4 to 20 mA current signal output.
Instrument has a standard RS232C interface
Specifications:
Measuring range: 0 ~ 10/100/1000 ppm/25% O2 (range automatic switching)
Optional 0 ~ 10/100/1000 ppm O2
Measurement accuracy <10 ppm at: +/- 0.5ppm range, <100 ppm pm: +/- 1ppm range, <1000 ppm pm: +/- 2ppm range, <25%: 0.1 +/- 1% range drift: =+/- 1% FS/7d range drift: =+/- 1% FS/7d
Alarm contact Capacity: AC 125V 0.3A, DC 110V 0.3A DC 30V 1A
Repeatability error: =1%
Response time: 10S 95%
Gas sampling: Membrane sampling pumps or built-in vacuum system (work needs 4 L / min flow stability)
Sample gas flow rate: 100 ~ 300 ml / min
Ambient temperature: 5 ‘C to +50 ‘C
Power source: 85 V-265V.AC
Frequency: 50 Hz +/- 5%
Weight: about 4 KG
Dimensions: L * W * D = 133mm * 270mm * 300mm