Pharmaceutical stability test chambers
XCH CSD Series Biomedical stability pharmaceutical chambers, Two case independent control,CSD series, A Chamber has three functions: temperature, humidity and illumination. near ultraviolet lamp tube can be optional,Meet the requirements of the stability test guidelines of the 2020 edition of pharmacopoeia; B Chamber can only control and record temperature and humidity.
Key Features:
• Temperature Control: Temperature Fluctuation ≤ ±0.5℃,Temperature Deviation ≤ ±1.0℃
• Humidity Control: Humidity Fluctuation ≤ ±3%RH,Humidity Deviation ≤ ±3%RH
• Structure Design: Adopt the latest airduct system design achieves good uniform of temperature and humidity form different parts inside the chamber.
• Insulation material: Integral polyurethane foam technology, with good heat preservation and moisture retention performance.
• Chamber Material: The liner is made of mirror stainless steel 304, no pollution source, easy to clean
• Refrigerating System: The imported fully enclosed industrial compressor, long-life, low noise
• Humidity System: Original imported capacitive humidity sensor, direct detection of humidity,maintenance-free
• Control System: Two sets color touch screen temperature and humidity controller,High control accuracy,With three levels permission,audit trail
• Safety device: compressor overheat and overpressure overload protection, Fan overheat protection、water shortage protection.
• Dual Protection: Equipped with double insurance dry fire protection system, equipped with independent overtemperature protection system, can automatically cut off the power supply
• Live Alarm: Temperature and humidity upper limit deviation alarm, independent overtemperature alarm.
• Illumination System: Equipped with illumination sensor,direct illumination display, equipped with illumination storage and printing system
• CSD series: A with function of illumination, Illumination Range: 100 ~ 8000Lux,Near Ultraviolet Range: 0.84 ~ 5w / m2 ,Illuminance deviation: 4500±500Lux
• Data Record: Equipped with data collection system conform to GMP requirements,Two sets Needle micro printer and USB output port as standard
• Meet a criterion: Meet ICHQ1A(R2)、Q1B、Chinese Pharmacopoeia 2020 edition
• Installed Power: AC220V±10% 50HZ
• Environment temperature: +5~35℃
• SMS alarm: Optional
Stability chamber specification:
MODEL | TEMP CONTROL RANGE(℃) | HUMIDITY CONTROL RANGE | CAPACITY(L) | INTER DIMEN W×D×H(mm) | EXTER DIMEN W×D×H(mm) | POWER(KW) | Shelf (Standard) | Shelf (Standard) |
XCH-320SD | 10 ~ 65 | 20 ~ 95% | 150 | 680×500×460 | 840×840×1880 | 2.5 | 2 | Up |
10 ~ 65 | 20 ~ 95% | 150 | 680×500×460 | 2.5 | 2 | Down | ||
XCH-520SD | 10 ~ 65 | 20 ~ 95% | 250 | 600×500×830 | 1560×870×1650 | 3.0 | 3 | Up |
10 ~ 65 | 20 ~ 95% | 250 | 600×500×830 | 3.0 | 3 | Down | ||
XCH-620SD | 10 ~ 65 | 20 ~ 95% | 300 | 600×500×1000 | 1560×870×1820 | 3.2 | 3 | Left |
10 ~ 65 | 20 ~ 95% | 300 | 600×500×1000 | 1560×870×1820 | 3 | Right | ||
Double chamber, independent control |
MODEL | TEMP CONTROL RANGE(℃) | HUMIDITY CONTROL RANGE | CAPACITY(L) | INTER DIMEN W×D×H(mm) | EXTER DIMEN W×D×H(mm) | POWER | Shelf (Standard) | Shelf (Standard) |
XCH-320CSD | 15 ~ 65 | 20 ~ 95% | 150 | 680×500×460 | 840×840×1880 | 2.5 | 2 | Up |
150 | 680×500×460 | 2.5 | 2 | Down | ||||
XCH-520CSD | 15 ~ 65 | 20 ~ 95% | 250 | 600×500×830 | 1560×870×1650 | 3.0 | 3 | Up |
250 | 600×500×830 | 1560×870×1650 | 3 | Down | ||||
XCH-620CSD | 15 ~ 65 | 20 ~ 95% | 300 | 600×500×1000 | 1560×870×1820 | 3.2 | 3 | Left |
600×500×1000 | 1560×870×1820 | 3 | Right | |||||
Double chamber, independent control |
Remark:
CSD series A equipped with temperature, humidity, visible light and NUV light four functions,Illumination measurement standard with visible and near ultraviolet sensors.Visible and near ultraviolet values are automatically printed and stored;Visible Range: 100 ~ 8000Lux,4500±500Lux,The total illumination of the light test shall not be less than 1.2×106 Lux·hr Near ultraviolet range: 0.84 ~ 5w/m2,Near ultraviolet energy shall not be less than 200w·hr/m2
Application:
XCH-SD series Medical Stability Testing Chamber Dual Chamber provide long-term stable temperature, humidity and light environment to detect the expiration date of drugs, used for long-term test, accelerated test, high humidity test, strong light irradiation test and drug stability test.
Advantages:
Sensitive, small system error
Two boxes of independent control
No pollution source and is easy to clean
Low cost