TPG-3030 Industrial Total Phosphorus Analyzer
The sample to be tested does not require any pretreatment. The water sample riser is directly inserted into the system water sample, and the total phosphorus concentration can be measured. The maximum measurement range of this equipment is 0.1~500mg/L TP. This method is mainly used for on-line automatic monitoring of total phosphorus concentration of waste (sewage) water discharge point source, surface water, etc.
Technical Indexes
1.Methods: National Standard GB11893-89 "Water quality - Determination of total phosphorus Ammonium molybdate spectrophotometric method".
2. Measuring range: 0-500mg/L TP (0-2mg/L;0.1-10mg/L;0.5-50mg/L; 1-100mg/L;5-500mg/L)
3. Accuracy: no more than±10% or no more than±0.2mg/L
4. Repeatability: no more than ±5% or no more than ±0.2 mg/L
5. Measurement period: minimum measuring period of 30 min, according to actual water samples, can be modified at 5 ~ 120min arbitrary digestion time.
6. Sampling period: the time interval (10 ~ 9999min adjustable) and the whole
point of measurement mode.
7. Calibration period: 1~99 days, any interval, any time adjustable.
8. Maintenance period: once a month, each about 30 min.
9. Reagent for value-based management:Less than 3 yuan/samples.
10. Output: RS-232;RS485;4~20mA three ways
11. Environmental requirement: temperature adjustable interior, it is
recommended temperature 5~28℃;humidity≤90%(no condensing)
12. Power supply: AC230±10%V, 50±10%Hz, 5A
13. Size: 1570 x500 x450mm(H*W*D).