Features: | ||||
1.With 3 1/2 LED, display is clear and precision is high. | ||||
2. Electrode constant compensation. | ||||
3.Manual temperature compensation. | ||||
4.It is equipped with DJS-1C platinum black probe. | ||||
Technical parameters: | ||||
1.Measuring range: The measuring range is (0~2×105) uS/cm. The instrument has five ranges and can be changed through band switch by manual (by electrode with constant). |
||||
Range | Measuring range | Resolution | ||
2 uS/cm | 0.001~2 uS/cm | 0.001 uS/cm | ||
20 uS/cm | 0.01~20 uS/cm | 0.01uS/cm | ||
200 uS/cm | 0.1~200 uS/cm | 0.1uS/cm | ||
2 mS/cm | 0.001~2.mS/cm | 0.001mS/cm | ||
20 mS/cm | 0.01~20 mS/cm | 0.01mS/cm | ||
NOTE: To measure high conductivity, a larger constant electrode is commonly used. If the conductivity is over 10000 uS/cm, the electrode with constant of 10 cm-1 is used and the measuring range is extended to 2×105 uS/cm. |
||||
2. Accuracy of the electronic unit : ±1.0%(FS) ±1bit | ||||
3.Accuracy of the meter: ±1.5%(FS) ±1bit | ||||
4. Range of Temp. compensation: 5-35℃ | ||||
5.Normal operation conditions: | ||||
(1) Ambient temperature : (5-35)℃ | ||||
(2) Relative humidity : not more than 85% | ||||
(3) Power source: 220V/110V | ||||
(4) Without any notable vibration. | ||||
(5)No strong electromagnetic interference to surrounding except the geomagnetic field. | ||||
Included Accessaries: | ||||
1.DJS-1C Platinum-black Probe | ||||
2.Multi-function Electrode Holder | ||||
3. Fuse BGXP Φ5*20 0.5A | ||||
4.Universal Adapter |