Scope
The instrument is in accordance with the People's Republic of China National Standard GB / T 261 - the design and manufacturing requirements, "Flash Point by Pensky Martens closed cup method" under
Features
1, the instrument is designed in full compliance with GB / T261 standards.
2 heating power continuous stepless power level by the voltage meter visual display, advanced and reasonable temperature control.
3, the structure is compact, stainless steel countertops and beautiful, easy to remove oil.
4, easy to operate, accurate test results, a number of measurement detection unit selected as the flash point of instrumentation.
Technical Parameters
1 Power supply: AC 220V ± 10%, 50Hz.
2, the heating device: heating power from 0 ~ 600W continuously adjustable.
3, the sample heating rate: 1 ~ 12 ℃ / min, controlled adjustable.
4, the electric mixing device: ⑴ 90 ~ 120 r / min for the test step A.
⑵ 250 ± 10 rev / min, applicable test procedure B.
6. Standard oil cup: ⑴ diameter: 50.7 ~ 50.8 mm.
⑵ Depth: 55.7 ~ 56.0 mm.
⑶ test oil capacity groove depth: 33.9 ~ 34.3 mm.
⑷ test oil capacity: about 70 ml.
7, ignition means: ⑴ ignition sources: gas (or other civilian combustible gas, the same below). ⑵ pilot aperture: 0.8 mm.
8, a thermometer: a bar mercury thermometer, in line with GB / T 261-2008 standards.
⑴ temperature range -5 ~ 110 ℃, indexing 0.5 ℃.
⑵ temperature range of 20 ~ 150 ℃, indexing 1.0 ℃.
⑶ temperature range of 90 ~ 370 ℃, dividing 2 ℃.
9, the ambient temperature: ≤35 ℃.
10, relative humidity: ≤85%.
11. Total power consumption: less than 650W.