Features
1.Comprehensive visible& audible alarms to make the infusion process safer and more reliable
2.Volume & rate memory function to keep the record of last time’s infusion
3.4.2” Large LCD screen for more clinical information to make reading easier and more clearer
4.Independent buttons&no changing data during working to avoid mis-operation and increase infusion safety
5.Ultrasound air detector, patented ultrasound detection for air in line
6.Infusion left time display
7.Body-weight infusion mode
8.Special lock to prevent Free-flow when door open
9.Optional function: 12V DC for ambulace and emergency use
10.Nurse all jack, RS232 for better satistaction of doctor’s clinical use
Technical Parameters
1.Infusion principle: Fingertip peristaltic Pump
2.I.V. set compatibility:All brands after calibration
3.Infusion mode:Drop mode, Rate mode, Body weight mode
4.Accuracy:+/-5%
5.Volume infused:0.1-9999.99ml
6.Infusion rate range:0.1-1200ml/h(0.1-100ml/h step by 0.1ml/h or 1ml/h, 100-1200ml/h step by 1ml/h)
7.Volume infused/Total volume infused:0-9999ml
8.Purge/Bolus rate: 200-1200ml/h, can be programmed by user, default rate 800ml/h
9.Bubble sensitivity:Detestability of air bubble>=40uL
10.Audible & Visible alarms: Air in line, occlusion, door open, VTBI completion, low battery, drop error, pause overtime, master CPU malfunction, monitoring CPU malfunction, battery exhausted, A.C. Disconnection, I.V. Set dislocation,Electric circuit jam, air detect circuit jam.
11.KVO:Rate>=10ml/h, KVO=3ml/h; Rate<10ml/h, KVO=1ml/h
12.Pause overtime alarm:1min 50s-2min
13.Power&battery:18VA; AC 100V-240V 50HZ/60HZ, DC 12V. Ni-MH battery can support more than 3 hours working at averate rate
14.Classification:Class II, type CF, IPX4
15.Dimension&weight:146mm(l)*146mm(w)*218mm(h), 2.5kg
16.Working environment:Temperature +5-+40c; relative humidity 20%-90%; atmospheric pressure, 70kpa-106kpa
17.Optional Function: DC input