Specifications xray dental machine
A. CE Mark
B. Wiress and Portable
C. High Quality
Features x ray equipment
Battery-pack
– Material : Lithium-Polymer
– Charge Condition
Current : 1.0 A
Voltage : 16.8 VDC
Temperature : 0 °C ~ 40° C
– Use condition
Max. current : 10 A
Voltage range : 12(min) ~ 16.8(MAX) V
Temperature : -20 °C ~ 60 °C
Charger
– Model : JBL910A-1600
– Input Voltage : 100 ~ 250 VAC
– Output Voltage : 16.8 VDC
– Current : 1.0A
Technical Data dental x ray machine
Name |
Rayme |
Tube Voltage |
70kV(fixed) |
Tube Current |
1mA(fixed) |
X-Ray Tube type |
Bipolar fixied X-ray tube |
X-Ray Tube model |
DG-073B-DC |
X-Ray Tube Focal Size |
0.3*0.3mm |
Cooling system |
Oil cooling |
Total filtration |
1.6 mm Al (Inherent filtration : 0.8mm, fixed added filter :0.7 mm Aluminum) |
Target angle |
21° |
Time Set-up function |
0.10 ~ 1.80 sec |
Source to image receptor distance |
>10cm (fixed by cone) |
X-ray field |
50 mm (round-style) |
Size of main body |
139(W) x 163(H) x 66.5mm(D) |
Weight of body |
About 2.3 kg |